A comprehensive investigation of the performance of a commercial scintillator system for applications in electron FLASH radiotherapy
Kevin Liu,Shannon Holmes,Emil Schüler,Sam Beddar
DOI: https://doi.org/10.1002/mp.17030
IF: 4.506
2024-03-22
Medical Physics
Abstract:Background Dosimetry in ultra‐high dose rate (UHDR) beamlines is significantly challenged by limitations in real‐time monitoring and accurate measurement of beam output, beam parameters, and delivered doses using conventional radiation detectors, which exhibit dependencies in ultra‐high dose‐rate (UHDR) and high dose‐per‐pulse (DPP) beamline conditions. Purpose In this study, we characterized the response of the Exradin W2 plastic scintillator (Standard Imaging, Inc.), a water‐equivalent detector that provides measurements with a time resolution of 100 Hz, to determine its feasibility for use in UHDR electron beamlines. Methods The W2 scintillator was exposed to an UHDR electron beam with different beam parameters by varying the pulse repetition frequency (PRF), pulse width (PW), and pulse amplitude settings of an electron UHDR linear accelerator system. The response of the W2 scintillator was evaluated as a function of the total integrated dose delivered, DPP, and mean and instantaneous dose rate. To account for detector radiation damage, the signal sensitivity (pC/Gy) of the W2 scintillator was measured and tracked as a function of dose history. Results The W2 scintillator demonstrated mean dose rate independence and linearity as a function of integrated dose and DPP for DPP ≤ 1.5 Gy (R2 > 0.99) and PRF ≤ 90 Hz. At DPP > 1.5 Gy, nonlinear behavior and signal saturation in the blue and green signals as a function of DPP, PRF, and integrated dose became apparent. In the absence of Cerenkov correction, the W2 scintillator exhibited PW dependence, even at DPP values 1.5 Gy, the W2 scintillator's response becomes nonlinear, likely due to limitations in the electrometer related to the high signal intensity.
radiology, nuclear medicine & medical imaging