Robust and Efficient Parametric Spectral Estimation in Atomic Force Microscopy

Bryan Yates,Aleksander Labuda,Martin Lysy
DOI: https://doi.org/10.48550/arXiv.1706.08938
2017-06-27
Applications
Abstract:An atomic force microscope (AFM) is capable of producing ultra-high resolution measurements of nanoscopic objects and forces. It is an indispensable tool for various scientific disciplines such as molecular engineering, solid-state physics, and cell biology. Prior to a given experiment, the AFM must be calibrated by fitting a spectral density model to baseline recordings. However, since AFM experiments typically collect large amounts of data, parameter estimation by maximum likelihood can be prohibitively expensive. Thus, practitioners routinely employ a much faster least-squares estimation method, at the cost of substantially reduced statistical efficiency. Additionally, AFM data is often contaminated by periodic electronic noise, to which parameter estimates are highly sensitive. This article proposes a two-stage estimator to address these issues. Preliminary parameter estimates are first obtained by a variance-stabilizing procedure, by which the simplicity of least-squares combines with the efficiency of maximum likelihood. A test for spectral periodicities then eliminates high-impact outliers, considerably and robustly protecting the second-stage estimator from the effects of electronic noise. Simulation and experimental results indicate that a two- to ten-fold reduction in mean squared error can be expected by applying our methodology.
What problem does this paper attempt to address?