Magnetocardiography on an isolated animal heart with a room-temperature optically pumped magnetometer

Kasper Jensen,Mark Alexander Skarsfeldt,Hans Stærkind,Jens Arnbak,Mikhail V. Balabas,Søren-Peter Olesen,Bo Hjorth Bentzen,Eugene S. Polzik
DOI: https://doi.org/10.1038/s41598-018-34535-z
IF: 4.6
2018-11-01
Scientific Reports
Abstract:Optically pumped magnetometers are becoming a promising alternative to cryogenically-cooled superconducting magnetometers for detecting and imaging biomagnetic fields. Magnetic field detection is a completely non-invasive method, which allows one to study the function of excitable human organs with a sensor placed outside the human body. For instance, magnetometers can be used to detect brain activity or to study the activity of the heart. We have developed a highly sensitive miniature optically pumped magnetometer based on cesium atomic vapor kept in a paraffin-coated glass container. The magnetometer is optimized for detection of biological signals and has high temporal and spatial resolution. It is operated at room- or human body temperature and can be placed in contact with or at a mm-distance from a biological object. With this magnetometer, we detected the heartbeat of an isolated guinea-pig heart, which is an animal widely used in biomedical studies. In our recordings of the magnetocardiogram, we can detect the P-wave, QRS-complex and T-wave associated with the cardiac cycle in real time. We also demonstrate that our device is capable of measuring the cardiac electrographic intervals, such as the RR- and QT-interval, and detecting drug-induced prolongation of the QT-interval, which is important for medical diagnostics.
multidisciplinary sciences
What problem does this paper attempt to address?