Expansion of measurement range: Tolerance analysis and application of the two-peak method for optical sensors with a few spectral peaks

Rui Yin,Lingxin Cao,Qingjie Huang,Lin Lu,Xu Yang,Jun Ying,Hongliang Yang,Wei Ji,Pengcheng Liu,Shouzhen Jiang
DOI: https://doi.org/10.1016/j.measurement.2023.112852
2023-04-13
Measurement: Journal of the International Measurement Confederation
Abstract:The two-peak method has been widely used to expand the measurement range of F-P cavities in the past decades. Previous studies have found that a dozen peak pairs should be involved to calculate the average value to reduce the possibility of misjudgment of the interference order m . This paper analyzes the impact of data error on the two-peak method and proposes the condition for correctly judging m using only one peak pair or peak-valley pair at the current technical level: m <1000. For experimental verification, a force sensor based on an integrated LiNbO 3 asymmetric Mach-Zehnder interferometer (AMZI) with m around 30 is fabricated and tested. The measurement range is originally 63 N due to the limitation of FSR (free spectral range). With the help of the two-peak method with only one peak-valley pair, the measurement range is successfully increased to 150 N. Experimental data show that the two-peak method exhibits a large tolerance when m is far less than 1000. This result extends the application of the two-peak method to the field of integrated optical sensors.
What problem does this paper attempt to address?