Late complete atrioventricular block after transcatheter perimembranous ventricular septal defect closure with the KONAR-MFTM VSD occluder: a case report and systematic review

Saïd Bichali,Ali Houeijeh,François Godart
DOI: https://doi.org/10.1017/s1047951124026891
2024-11-01
Cardiology in the Young
Abstract:The soft nitinol KONAR-MFTM ventricular septal defect (VSD) Multifunctional Occluder (MFO) device is increasingly used for transcatheter perimembranous ventricular septal defect closure. We report for the first time a case of delayed complete atrioventricular block with pacemaker implantation 20 months post-procedure. Through a systematic review, the overall rate of persistent complete atrioventricular block was 0.6% with this device, but follow-up duration was limited.
pediatrics,cardiac & cardiovascular systems
What problem does this paper attempt to address?