Hemodynamic Evaluation of Type B Aortic Dissection Using Compressed Sensing Accelerated 4D Flow MRI

Ozden Kilinc,Stanley Chu,Justin Baraboo,Elizabeth K. Weiss,Joshua Engel,Anthony Maroun,Daniel Giese,Ning Jin,Kelvin Chow,Xiaoming Bi,Rachel Davids,Christopher Mehta,S. Chris Malaisrie,Andrew Hoel,James Carr,Michael Markl,Bradley D. Allen
DOI: https://doi.org/10.1002/jmri.28432
IF: 4.4
2022-09-25
Journal of Magnetic Resonance Imaging
Abstract:Background 4D Flow MRI is a quantitative imaging technique to evaluate blood flow patterns; however, it is unclear how compressed sensing (CS) acceleration would impact aortic hemodynamic quantification in type B aortic dissection (TBAD). Purpose To investigate CS‐accelerated 4D Flow MRI performance compared to GRAPP‐accelerated 4D Flow MRI (GRAPPA) to evaluate aortic hemodynamics in TBAD. Study Type Prospective. Population Twelve TBAD patients, two volunteers. Field Strength/Sequence 1.5T, 3D time‐resolved cine phase‐contrast gradient echo sequence. Assessment GRAPPA (acceleration factor [R] = 2) and two CS‐accelerated (R = 7.7 [CS7.7] and 10.2 [CS10.2]) 4D Flow MRI scans were acquired twice for interscan reproducibility assessment. Voxelwise kinetic energy (KE), peak velocity (PV), forward flow (FF), reverse flow (RF), and stasis were calculated. Plane‐based mid‐lumen flows were quantified. Imaging times were recorded. Tests Repeated measures analysis of variance, Pearson correlation coefficients (r), intraclass correlation coefficients (ICC). P
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?