Usefulness of a lead-acrylic shield for reducing lens dose of assistant in x-ray CT examination

Hajime Ito,Kosuke Matsubara,Ikuo Kobayashi,Takayuki Sakai,Tomonori Isobe,Noriyuki Yanagawa,Shigehiro Ochi
DOI: https://doi.org/10.1088/1361-6498/ac7919
2022-06-16
Journal of Radiological Protection
Abstract:In CT examinations, the usefulness of protective glasses for reducing lens exposure to assistants has been reported. The present study aimed to compare the dose reduction effect for assistants with lead-acrylic shields and protective glasses (0.07 mmPb, 0.5 mmPb) during CT examination. The air dose distribution in a CT examination room with and without a lead-acrylic shield was compared. It was found that the amount of scattered radiation was significantly reduced by installing a lead-acrylic shield at the CT gantry aperture. Moreover, the reduction rate of air kerma at the assistant's lens was higher using the lead acrylic shield than with the protective glasses—95.7% during head holding and 76.1% during assisted ventilation.
environmental sciences,public, environmental & occupational health,radiology, nuclear medicine & medical imaging,nuclear science & technology
What problem does this paper attempt to address?