Coating of Sandblasted and Acid-Etched Dental Implants With Tantalum Using Vacuum Plasma Spraying

Xian Zhou,Xiulian Hu,Ye Lin,Hu, Xiulian,Lin, Ye,Zhou, Xian
DOI: https://doi.org/10.1097/ID.0000000000000727
2019-10-03
Implant Dentistry
Abstract:Purpose: The objective was to prepare tantalum (Ta)-coated sandblasted and acid-etched (SLA) dental implants using vacuum plasma spraying (VPS) and to analyze their morphologies. Materials and Methods: Twelve SLA implants were coated with Ta using VPS. The topographies of the coatings and Ta/SLA surface interfaces were examined using scanning electron microscopy. The thickness at 4 locations for 6 Ta-coated and 6 uncoated SLA implants and pore sizes of the neck, central, and root areas of Ta-coated implants were measured. SPSS v20.0 was used for statistical analysis. Results: The Ta coatings were rough and consisted of pitted structures with various pore sizes; no cracks were observed. The Ta/SLA surface interface was tightly bonded. The 95% confidence interval of the Ta coating thickness was (114.0759, 129.3574). The maximal pore diameter was concentrated at 200 to 400 nm. Conclusion: SLA dental implants were successfully coated with Ta using VPS. The nanoporous structure of these implants may facilitate osseointegration compared with uncoated SLA implants.
dentistry, oral surgery & medicine
What problem does this paper attempt to address?