Application of a novel diamond detector for commissioning of FLASH radiotherapy electron beams

Gianluca Verona Rinati,Giuseppe Felici,Federica Galante,Alessia Gasparini,Rafael Kranzer,Giulia Mariani,Matteo Pacitti,Giuseppe Prestopino,Andreas Schüller,Verdi Vanreusel,Dirk Verellen,Claudio Verona,Marco Marinelli
DOI: https://doi.org/10.1002/mp.15782
IF: 4.506
2022-06-03
Medical Physics
Abstract:Purpose A diamond detector prototype was recently proposed by Marinelli et al (Medical Physics 2022, https://doi.org/10.1002/mp.15473) for applications in Ultra‐High‐Dose‐Per‐Pulse (UH‐DPP) and Ultra‐High‐Dose‐Rate (UH‐DR) beams, as used in FLASH radiotherapy. In the present study, such so‐called flashDiamond (fD) was investigated from the dosimetric point of view, under pulsed electron beam irradiation. It was then used for commissioning of an ElectronFlash linac (SIT S.p.A, Italy) both in conventional and UH‐DPP modalities. Methods Detector calibration was performed in reference conditions, under 60Co and electron beam irradiation. Its response linearity was investigated in UH‐DPP conditions. For this purpose the DPP was varied in the 1.2 – 11.9 Gy range, by changing either the beam applicator or the pulse duration from 1 μs to 4 μs. A dosimetric validation of the fD detector prototype was then performed in conventional modality, by measuring Percentage Depth Dose (PDD) curves, beam profiles and Output Factors (OFs). All such measurements were carried out in a motorized water phantom. The obtained results were compared with the ones from commercially available dosimeters, namely a microDiamond, an Advanced Markus ionization chamber, a Silicon Diode detector and EBT‐XD GAFchromic films. Finally, the fD detector was used to fully characterize the 7 MeV and 9 MeV UH‐DPP electron beams delivered by the ElectronFlash linac. In particular, PDDs, beam profiles and OFs were measured, for both energies and all the applicators, and compared with the ones from EBT‐XD films irradiated in the same experimental conditions. Results The fD calibration coefficient resulted to be independent from the investigated beam qualities. The detector response was found to be linear in the whole investigated DPP range. A very good agreement was observed among PDDs, beam profiles and OFs measured by the fD prototype and reference detectors, both in conventional and UH‐DPP irradiation modalities. Conclusions The fD detector prototype was validated from the dosimetric point of view against several commercial dosimeters in conventional beams. It was proved to be suitable in UH‐DPP and UH‐DR conditions, for which no other commercial real‐time active detector is available to date. It was shown to be a very useful tool to perform fast and reproducible beam characterizations in standard clinical motorized water phantom setups. All of the above demonstrates the suitability of the proposed detector for commissioning of UH‐DR linac beams for preclinical FLASH radiotherapy applications. This article is protected by copyright. All rights reserved
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?