Comparison of Automated Brain Volume Measures obtained with NeuroQuant® and FreeSurfer

Alfred L Ochs,David E Ross,Megan D Zannoni,Tracy J Abildskov,Erin D Bigler,Alfred L. Ochs,David E. Ross,Megan D. Zannoni,Tracy J. Abildskov,Erin D. Bigler,
DOI: https://doi.org/10.1111/jon.12229
2015-02-26
Journal of Neuroimaging
Abstract:PURPOSE: To examine intermethod reliabilities and differences between FreeSurfer and the FDA-cleared congener, NeuroQuant, both fully automated methods for structural brain MRI measurements.MATERIALS AND METHODS: MRI scans from 20 normal control subjects, 20 Alzheimer's disease patients, and 20 mild traumatically brain-injured patients were analyzed with NeuroQuant and with FreeSurfer. Intermethod reliability was evaluated.RESULTS: Pairwise correlation coefficients, intraclass correlation coefficients, and effect size differences were computed. NeuroQuant versus FreeSurfer measures showed excellent to good intermethod reliability for the 21 regions evaluated (r: .63 to .99/ICC: .62 to .99/ES: -.33 to 2.08) except for the pallidum (r/ICC/ES = .31/.29/-2.2) and cerebellar white matter (r/ICC/ES = .31/.31/.08). Volumes reported by NeuroQuant were generally larger than those reported by FreeSurfer with the whole brain parenchyma volume reported by NeuroQuant 6.50% larger than the volume reported by FreeSurfer. There was no systematic difference in results between the 3 subgroups.CONCLUSION: NeuroQuant and FreeSurfer showed good to excellent intermethod reliability in volumetric measurements for all brain regions examined with the only exceptions being the pallidum and cerebellar white matter. This finding was robust for normal individuals, patients with Alzheimer's disease, and patients with mild traumatic brain injury.
radiology, nuclear medicine & medical imaging,clinical neurology,neuroimaging
What problem does this paper attempt to address?