Anatomical considerations for the design and applications of cochlear implants

Helge E. Rask-Andersen,Wei Liu,Annelies Schrott-Fischer,Rudolf Glueckert
DOI: https://doi.org/10.1121/1.4920386
2015-04-01
The Journal of the Acoustical Society of America
Abstract:This presentation aims to display some important anatomical characteristics and variations of the human cochlea that may influence outcomes with cochlear implants. I will describe the complexity of the “hook” region and how it may challenge atraumatic insertion of an electrode array. In addition, I will discuss advantages and disadvantages of various cochleostomy approaches and of various trajectories for insertions. These anatomical considerations are informed by micro dissections of human temporal bones and plastic casts of different parts of the bones. In addition, the studied microanatomy of the basilar membrane may inform insertions and insertion depths needed to preserve cochlear structures. In particular, the studies suggest that the vulnerability of the membrane may be greatest at the apex and this possibility has implications for electrode designs and insertions. Also, the apical region may be most susceptible to inflammation and fibrosis due to its anatomy. The anatomical considerations for safe insertions of electrodes, including preservation of the remaining neural tissue, will be discussed. This work was supported by grants from Uppsala University Hospital and Uppsala University; the Tysta Skolan Foundation; the Swedish Deafness Foundation; the European Community 7th Framework Programme; and kind private donations from Börje Runögård.
acoustics,audiology & speech-language pathology
What problem does this paper attempt to address?