Towards inferring positioning of straight cochlear‐implant electrode arrays during insertion using real‐time impedance sensing

Katherine E. Riojas,Trevor L. Bruns,Josephine Granna,Miriam R. Smetak,Robert F. Labadie,Robert J. Webster
DOI: https://doi.org/10.1002/rcs.2609
2024-01-05
International Journal of Medical Robotics and Computer Assisted Surgery
Abstract:Abstract Background Cochlear‐implant electrode arrays (EAs) are currently inserted with limited feedback, and impedance sensing has recently shown promise for EA localisation. Methods We investigate the use of impedance sensing to infer the progression of an EA during insertion. Results We show that the access resistance component of bipolar impedance sensing can detect when a straight EA reaches key anatomical locations in a plastic cochlea and when each electrode contact enters/exits the cochlea. We also demonstrate that dual‐sided electrode contacts can provide useful proximity information and show the real‐time relationship between impedance and wall proximity in a cadaveric cochlea for the first time. Conclusion The access resistance component of bipolar impedance sensing has high potential for estimating positioning information of EAs relative to anatomy during insertion. Main limitations of this work include using saline as a surrogate for human perilymph in ex vivo models and using only one type of EA.
surgery
What problem does this paper attempt to address?