Validation of high-resolution gamma-ray computed tomography for quantitative gas holdup measurements in centrifugal pumps

André Bieberle,Thomas Schäfer,Martin Neumann,Uwe Hampel
DOI: https://doi.org/10.1088/0957-0233/26/9/095304
IF: 2.398
2015-07-30
Measurement Science and Technology
Abstract:In this article, the capability of high-resolution gamma-ray computed tomography (HireCT) for quantitative gas–liquid phase distribution measurements in commercially available industrial pumps is experimentally investigated. The object of interest thereby operates under two-phase flow conditions. HireCT System comprises a collimated 137Cs isotopic source, a radiation detector arc with a multi-channel signal processing unit, and a rotary unit enabling CT scans of objects with diameters of up to 700 mm. The accuracy of gas holdup measurements was validated on a sophisticated modular test mockup replicating defined gas–liquid distributions, which are expected in impeller chambers of industrial centrifugal pumps under two-phase operation. Stationary as well as rotation-synchronized CT scanning techniques have been analyzed, which are both used to obtain sharply resolved gas phase distributions in rotating structures as well as non-rotating zones. A measuring accuracy of better than 1% absolute for variously distributed static gas holdups in the rotating frame has been verified with the modular test mockup using HireCT.