Influence of PEEK Coating on Hip Implant Stress Shielding: A Finite Element Analysis

Jesica Anguiano-Sanchez,Oscar Martinez-Romero,Hector R. Siller,Jose A. Diaz-Elizondo,Eduardo Flores-Villalba,Ciro A. Rodriguez
DOI: https://doi.org/10.1155/2016/6183679
IF: 2.809
2016-01-01
Computational and Mathematical Methods in Medicine
Abstract:Stress shielding is a well-known failure factor in hip implants. This work proposes a design concept for hip implants, using a combination of metallic stem with a polymer coating (polyether ether ketone (PEEK)). The proposed design concept is simulated using titanium alloy stems and PEEK coatings with thicknesses varying from 100 to 400 μ m. The Finite Element analysis of the cancellous bone surrounding the implant shows promising results. The effective von Mises stress increases between 81 and 92% for the complete volume of cancellous bone. When focusing on the proximal zone of the implant, the increased stress transmission to the cancellous bone reaches between 47 and 60%. This increment in load transferred to the bone can influence mineral bone loss due to stress shielding, minimizing such effect, and thus prolonging implant lifespan.
mathematical & computational biology