Isomer separation of polybrominated diphenyl ether metabolites using nanoESI-TIMS-MS

Kendra J. Adams,Dennise Montero,Diana Aga,Francisco Fernandez-Lima
DOI: https://doi.org/10.1007/s12127-016-0198-z
2016-04-21
International Journal for Ion Mobility Spectrometry
Abstract:In this paper, high-resolution nano-electrospray ionization-trapped ion mobility spectrometry coupled to mass spectrometry (nESI-TIMS-MS) is used for the study of hydroxylated polybrominated diphenyl ether (OH-PBDE) metabolites. In particular, experimental ion-neutral collision cross sections (CCS) were measured for five structural OH-PBDE isomers using TIMS-MS. Candidate structures were proposed for each IMS band observed in good agreement with the experimental CCS measurements (5 % error). The analytical power of TIMS-MS to baseline and partially separate structural isomers of OH-BDE in binary and ternary mixtures is shown for single charge species with a mobility resolving power of RIMS ~ 400. This work provides the proof of concept for the analysis of low concentration OH-PBDE in environmental samples based on accurate collision cross section and mass measurements without the need for derivatization and pre-fractionation protocols, thus significantly reducing the cost and analysis time.
What problem does this paper attempt to address?