Novel approach for overcoming effects of diffraction in measurements of the nonlinear B/A parameter in liquidlike media

Kirk D. Wallace,Mark R. Holland,James G. Miller
DOI: https://doi.org/10.1121/1.4785245
2004-10-01
The Journal of the Acoustical Society of America
Abstract:Finite amplitude measurements of the nonlinear B/A parameter in a liquidlike medium such as soft tissue can be complicated by the diffraction pattern associated with a bounded ultrasonic source. The goal of this work is to design and implement an approach based on the use of a stainless steel delay-line to position the specimen of interest either at the last axial maximum or in the far-field portion of the diffraction pattern from a piston source. A Burgers equation enhanced, nonlinear angular spectrum simulation was used to design the experimental apparatus by modeling the ultrasonic fields. A 0.6-mm-diam hydrophone receiver was mechanically scanned in a water-bath to measure the finite amplitude ultrasonic fields emerging from the delay-lines. A thin-plate tungsten resonator was used to suppress the fundamental relative to the second harmonic. As evidence of the effectiveness of this approach for overcoming complications of data analysis arising from diffraction, the nonlinear parameter B/A, the small-signal speed of sound, and the attenuation coefficient were successfully measured over a wide range of isopropanol-water mixtures. Measurements of B/A had an average uncertainty of less than 5% over the range of values associated with biological tissues (4.6 to 11.9). [Work supported by NIH R01HL72761.]