Optical diffraction for measurements of nano-mechanical bending

Rodolfo I. Hermans,Benjamin Dueck,Joseph Wafula Ndieyira,Rachel A. McKendry,Gabriel Aeppli
DOI: https://doi.org/10.1038/srep26690
2015-09-01
Abstract:Micromechanical transducers such as cantilevers for AFM often rely on optical readout methods that require illumination of a specific region of the microstructure. Here we explore and exploit the diffraction effects that have been previously neglected when modeling cantilever bending measurement techniques. The illumination of a cantilever end causes an asymmetric diffraction pattern at the photodetector that significantly affects the calibration of the signal in the popular optical beam deflection technique (OBDT). Conditions for optimized linear signals that avoid detection artifacts conflict with small numerical aperture illumination and narrow cantilevers which are softer and therefore more sensitive. Embracing diffraction patterns as a physical measurable allows a richer detection technique that decouples measurements of tilt and curvature and simultaneously relaxes the requirements on the alignment of illumination and detector. We show analytical results, numerical simulations and physiologically relevant experimental data demonstrating the usefulness of these diffraction features. We offer experimental design guidelines and identify and quantify possible sources of systematic error of up to 10% in OBDT. We demonstrate a new nanometre resolution detection method that can replace OBDT, where Frauenhofer and Bragg diffraction effects from finite sized and patterned cantilevers are exploited. Such effects are readily generalized to arrays, and allow transmission detection of mechanical curvature, enabling in-line instruments. In particular, a cantilever with a periodic array of slots produces Bragg peaks which can be analyzed to deduce the cantilever curvature. We highlight the comparative advantages over OBDT by detecting molecular activity of antibiotic Vancomycin, with an RMS noise equivalent to less than $2.5 \mu M$ (1.5 nm), as example of possible multi-maker bio-assays.
Instrumentation and Detectors
What problem does this paper attempt to address?
The problem that this paper attempts to solve is: the limitations and error problems existing in the traditional optical beam deflection technique (OBDT) when measuring micro - cantilever bending. Specifically, when using OBDT for measurement, it is necessary to precisely align the laser spot, and the spot size must be much smaller than the cantilever size to avoid the influence of edge diffraction effects. However, this requirement limits the versatility and miniaturization applications of the technology, especially in non - professional laboratory environments. In addition, OBDT is very sensitive to small perturbations in the optical path, such as particles in the liquid or temperature gradients, etc., which will all affect the accuracy of the measurement. For this reason, the author proposes a new method based on far - field diffraction imaging to replace the traditional OBDT. This method utilizes the diffraction pattern generated when the micro - cantilever bends, which can not only measure the tilt and curvature change of the cantilever more accurately, but also greatly relax the requirements for the position of the laser spot and the detector position, making the technology more robust and suitable for a wider range of scenarios, including large - scale multiplex detection in non - professional laboratory environments. Through theoretical analysis, numerical simulation and experimental verification, the author shows the effectiveness and advantages of the new method, especially showing the same results as OBDT in detecting the binding reaction between the antibiotic vancomycin and the target peptide, proving its potential application value in biomedical detection.