A sensitive sandwich-type electrochemical immunosensor using nitrogen-doped graphene/metal-organic framework-derived CuMnCoOx and Au/MXene for the detection of breast cancer biomarker
Rajalakshmi Sakthivel,Yu-Chien Lin,Min-Chin Yu,Udesh Dhawan,Xinke Liu,Jung-Chih Chen,Ching-Wei Tung,Ren-Jei Chung
DOI: https://doi.org/10.1016/j.colsurfb.2024.113755
Abstract:In terms of cancer-related deaths among women, breast cancer (BC) is the most common. Clinically, human epidermal growth receptor 2 (HER2) is one of the most commonly used diagnostic biomarkers for facilitating BC cell proliferation and malignant growth. In this study, a disposable gold electrode (DGE) modified with gold nanoparticle-decorated Ti3C2Tx (Au/MXene) was utilized as a sensing platform to immobilize the capturing antibody (Ab1/Au/MXene). Subsequently, nitrogen-doped graphene (NG) with a metal-organic framework (MOF)-derived copper-manganese-cobalt oxide, tagged as NG/CuMnCoOx, was used as a probe to label the detection antibody (Ab2). A sandwich-type immunosensor (NG/CuMnCoOx/Ab2/HER2-ECD /Ab1/Au/MXene/DGE) was developed to quantify HER2-ECD. NG/CuMnCoOx enhances the conductivity, electrocatalytic active sites, and surface area to immobilize Ab2. In addition, Au/MXene facilitates electron transport and captures more Ab1 on its surface. Under optimal conditions, the resultant immunosensor displayed an excellent linear range of 0.0001 to 50.0 ng. mL-1. The detection limit was 0.757 pg·mL-1 with excellent selectivity, appreciable reproducibility, and high stability. Moreover, the applicability for determining HER2-ECD in human serum samples indicates its ability to monitor tumor markers clinically.