Sensitivity Improvement of a MXene- Immobilized SPR Sensor With Ga-Doped-ZnO for Biomolecules Detection

Rajeev Kumar,Sarika Pal,Yogendra Kumar Prajapati,Santosh Kumar,J. P. Saini
DOI: https://doi.org/10.1109/jsen.2022.3154099
IF: 4.3
2022-04-01
IEEE Sensors Journal
Abstract:This work presents a surface plasmon resonance (SPR) sensor based on Gallium (Ga)-doped-Zinc oxide (ZnO) and Ti3 C2 Tx (MXene) for sensitivity enhancement. The smaller resistivity and high conductivity of the pulsed layer deposited Ga-doped-ZnO are used to improve the sensitivity up to (264.59 °/RIU). The detection accuracy (0.1152°) and figure of merit (30.48 RIU−1) are also evaluated for the proposed sensor. The effects of using different prisms (SiO2, BaF2, CsF, and BK7), plasmonic metals (Au, Ag, Cu, and Al) and doping in ZnO (intermediate layer) on the performance parameters of the proposed SPR sensor are investigated for comparative analysis. The comparative performance analysis presents the maximum sensitivity for the SPR sensor structure: Prism SiO2/Au/Ga-doped-ZnO/MXene. The field distribution at different layer interfaces of the SPR sensor configurations has also been analyzed using FEM simulations. This work is suitable for the detection of biomolecules and biochemicals
engineering, electrical & electronic,instruments & instrumentation,physics, applied
What problem does this paper attempt to address?