Long term measurement of the $^{222}$Rn concentration in the Canfranc Underground Laboratory
J. Amaré,I. Bandac,A. Blancas,S. Borjabad,S. Buisán,S. Cebrián,D. Cintas,I. Coarasa,E. García,M. Martínez,R. Núñez Lagos,M.A. Oliván,Y. Ortigoza,A. Ortiz de Solórzano,C. Pérez,J. Puimedón,S. Rodríguez,A. Salinas,M.L. Sarsa,P. Villar
DOI: https://doi.org/10.1140/epjc/s10052-022-10859-z
2022-03-26
Abstract:We report the results of six years (2013-2018) of measurements of $^{222}$Rn air concentration, relative humidity, atmospheric pressure and temperature in the halls A, B and C of the Canfranc Underground Laboratory (LSC). We have calculated all the Pearson correlation coefficients among these parameters and we have found a positive correlation between the $^{222}$Rn concentration and the relative humidity. Both correlated variables show a seasonal periodicity. The joint analysis of laboratory data and four years (2015-2018) of the meteorological variables outside the laboratory shows the correlation between the $^{222}$Rn concentration and the outside temperature. The collected information stresses the relevance of designing good Rn-mitigation strategies in current and future experiments at LSC; in particular, we have checked for two years (2017-2018) the good performance of the mitigation procedure of the ANAIS--112 experiment. Finally, in another measurement (2019-2021) for two years of live time, we report an upper limit to the residual $^{222}$Rn content of the radon-free air provided by the radon abatement system installed in the laboratory.
Instrumentation and Detectors,Instrumentation and Methods for Astrophysics