Inductance Calculation of Multilayer Circular Printed Spiral Coils

Tao Liu,Zhiqiang Wei,Haokun Chi,Bo Yin
DOI: https://doi.org/10.1088/1742-6596/1176/6/062045
2019-03-01
Journal of Physics: Conference Series
Abstract:In the past few years, implanted biomedical devices have been widely used in healthy and medical applications and wireless power transfer (WPT) techniques used in these devices have been become a research hot topic. The printed spiral coils (PSCs), which have characteristics of small volume, low cost, high precision and high-integration, are better suited for WPT system. It is noteworthy that mutual inductance value of PSCs is a significant parameter which can influence the transfer efficiency of WPT system. Considering the geometry of multilayer circular PSCs and the distance between two adjacent layers, the previous methods are not suitable for calculating the mutual inductance of multilayer circular PSCs. In this study, an approximate calculation formula of mutual inductance between two single-layer circular PSCs is presented based on the existing research. On this basis, the self-inductance values of double-layer circular PSCs are calculated by MATLAB, the calculation results are verified by the ANSYS Q3D simulation and measurement results.
What problem does this paper attempt to address?