Medical Applications of a Small Sample X-Ray Fluorescence System

A. C. Alfrey,L. L. Nunnelley,H. Rudolph,W. R. Smythe
DOI: https://doi.org/10.1154/s0376030800008041
1975-01-01
Abstract:A quantitative x-ray fluorescence (XRF) analytical system has been developed with the emphasis on small sample size. The system employs a two kilowatt tungsten anode x-ray tube, and an 80 mm 2 lithium drifted silicon x-ray detector arranged in a compact geometry. Operation at 55 kV and 20 m A give a sensitivity of 153 counts/min/ng at Z=42 (molybdenum). A computer controlled target changer makes it capable of analyzing 32 targets automatically. The sample target is rotated during irradiation, so that the average sample position is defined to 0.2 mm. Several sets of 5 to 9 calibration targets have exhibited a standard deviation of less than 1%.