NaI gamma camera performance for high energies: Effects of crystal thickness, photomultiplier tube geometry and light guide thickness
Valerio Cosmi,Beien Wang,Marlies C. Goorden,Freek J. Beekman
DOI: https://doi.org/10.1002/mp.17043
IF: 4.506
2024-04-05
Medical Physics
Abstract:Background Gamma camera imaging, including single photon emission computed tomography (SPECT), is crucial for research, diagnostics, and radionuclide therapy. Gamma cameras are predominantly based on arrays of photon multipliers tubes (PMTs) that read out NaI(Tl) scintillation crystals. In this way, standard gamma cameras can localize ɣ‐rays with energies typically ranging from 30 to 360 keV. In the last decade, there has been an increasing interest towards gamma imaging outside this conventional clinical energy range, for example, for theragnostic applications and preclinical multi‐isotope positron emission tomography (PET) and PET‐SPECT. However, standard gamma cameras are typically equipped with 9.5 mm thick NaI(Tl) crystals which can result in limited sensitivity for these higher energies. Purpose Here we investigate to what extent thicker scintillators can improve the photopeak sensitivity for higher energy isotopes while attempting to maintain spatial resolution. Methods Using Monte Carlo simulations, we analyzed multiple PMT‐based configurations of gamma detectors with monolithic NaI (Tl) crystals of 20 and 40 mm thickness. Optimized light guide thickness together with 2‐inch round, 3‐inch round, 60 × 60 mm2 square, and 76 × 76 mm2 square PMTs were tested. For each setup, we assessed photopeak sensitivity, energy resolution, spatial, and depth‐of‐interaction (DoI) resolution for conventional (140 keV) and high (511 keV) energy ɣ using a maximum‐likelihood algorithm. These metrics were compared to those of a "standard" 9.5 mm‐thick crystal detector with 3‐inch round PMTs. Results Estimated photopeak sensitivities for 511 keV were 27% and 53% for 20 and 40 mm thick scintillators, which is respectively, 2.2 and 4.4 times higher than for 9.5 mm thickness. In most cases, energy resolution benefits from using square PMTs instead of round ones, regardless of their size. Lateral and DoI spatial resolution are best for smaller PMTs (2‐inch round and 60 × 60 mm2 square) which outperform the more cost‐effective larger PMT setups (3‐inch round and 76 × 76 mm2 square), while PMT layout and shape have negligible (
radiology, nuclear medicine & medical imaging