Evaluation and improvement the safety of total marrow irradiation with helical tomotherapy using repeat failure mode and effects analysis

Jiuling Shen,Xiaoyong Wang,Di Deng,Jian Gong,Kang Tan,Hongli Zhao,Zhirong Bao,Jinping Xiao,An Liu,Yunfeng Zhou,Hui Liu,Conghua Xie
DOI: https://doi.org/10.1186/s13014-019-1433-7
IF: 4.309
2019-12-01
Radiation Oncology
Abstract:Abstract Background & purpose Helical tomotherapy has been applied to total marrow irradiation (HT-TMI). Our objective was to apply failure mode and effects analysis (FMEA) two times separated by 1 year to evaluate and improve the safety of HT-TMI. Materials and methods A multidisciplinary team was created. FMEA consists of 4 main steps: (1) Creation of a process map; (2) Identification of all potential failure mode (FM) in the process; (3) Evaluation of the occurrence (O), detectability (D) and severity of impact (S) of each FM according to a scoring criteria (1–10), with the subsequent calculation of the risk priority number (RPN=O*D*S) and (4) Identification of the feasible and effective quality control (QC) methods for the highest risks. A second FMEA was performed for the high-risk FMs based on the same risk analysis team in 1 year later. Results A total of 39 subprocesses and 122 FMs were derived. First time RPN ranged from 3 to 264.3. Twenty-five FMs were defined as being high-risk, with the top 5 FMs (first RPN/ second RPN): (1) treatment couch movement failure (264.3/102.8); (2) section plan dose junction error in delivery (236.7/110.4); (3) setup check by megavoltage computed tomography (MVCT) failure (216.8/94.6); (4) patient immobilization error (212.5/90.2) and (5) treatment interruption (204.8/134.2). A total of 20 staff members participated in the study. The second RPN value of the top 5 high-risk FMs were all decreased. Conclusion QC interventions were implemented based on the FMEA results. HT-TMI specific treatment couch tests; the arms immobilization methods and strategy of section plan dose junction in delivery were proved to be effective in the improvement of the safety.
oncology,radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?