Research on scene temperature calibration method for medical infrared temperature measurement components

Xinsai Wang,Xiaoer Feng,Mingming Li,Weikai Fan
DOI: https://doi.org/10.1117/12.2541793
2020-02-14
Abstract:Due to its unique advantages, infrared temperature measurement technology has been used more and more widely in medical auxiliary diagnosis, and the change of ambient temperature is an important factor affecting medical infrared temperature measurement components. This paper builds a simulation platform for ambient temperature. At a fixed distance, by changing the ambient temperature and the target temperature, a large amount of infrared temperature measurement data is collected, and a two-dimensional compensation model of the scene temperature variation is obtained by combining the infrared focal plane non-uniformity correction and the least squares fitting. The actual scene experiment proves that the temperature measurement accuracy and stability have achieved satisfactory results.
What problem does this paper attempt to address?