Development and performance testing of the low-cost, 3D-printed, smartphone-compatible ‘Tansen Videolaryngoscope’ vs. Pentax-AWS videolaryngoscope vs. direct Macintosh laryngoscope

Christopher T Lambert,Stephen C John,Anna V John,Christopher T. Lambert,Stephen C. John,Anna V. John
DOI: https://doi.org/10.1097/eja.0000000000001264
2020-07-15
European Journal of Anaesthesiology
Abstract:While videolaryngoscopes help in the management of difficult airways, they remain too expensive for those with limited resources. We have developed a robust, re-usable, low-cost videolaryngoscope at United Mission Hospital Tansen, Nepal, by combining a smartphone-compatible endoscope capable of capturing still and video images with a three dimensional-printed, channelled, hyperangulated blade. The computer-aided design file for the videolaryngoscope blade was emailed and printed in London before evaluation of its performance on a difficult airway manikin.
anesthesiology
What problem does this paper attempt to address?