Quality control of cardiopulmonary exercise equipment

Tjeu Souren,Erwin De Bliek,Johan Roeykens,Kevin De Soomer,Ellie Oostveen
DOI: https://doi.org/10.1183/13993003.congress-2020.3786
2020-09-07
Abstract:<p><b>Introduction:</b> Quality control of cardiopulmonary exercise equipment in daily clinical practice is usually restricted to the manufacturer's recommendations for calibration and maintenance. It is uncommon to perform an overall check on the accuracy of the measurements of ventilation (V'E), O2 uptake (V'O2) and CO2 production (V'CO2). Metabolic simulators have been developed specifically for this purpose.</p><p><b>Aim:</b> To verify the performance of 4 different exercise testing devices from 4 manufacturers employed at different departments in an university hospital setting.</p><p><b>Methods:</b> A metabolic simulator (Relitech Systems, Netherlands), delivered stroke volumes of 2 Liters exactly at 4 frequencies (a: 20, b: 40, c: 60, d: 80 strokes/min), with associated values of both V'O2 and V'CO2 of 1, 2, 3 and 4 L/min, respectively. Most common, a 3% measurement error is acceptable.</p><p><b>Results:</b> The relative difference between measured and delivered values of V'E, VO2 and V'CO2 of the devices are presented in the table.</p><p><b>Discussion and Conclusion:</b> The metabolic simulator clearly identified devices that did not reliably measure V'E, V'O2 and/or V'CO2. We recommend verification with a metabolic simulator to become a routine part of the yearly quality check of exercise equipment.</p>