Dosimetric calibration of anatomy‐specific ultra‐high dose rate electron irradiation platform for preclinical FLASH radiobiology experiments
Jinghui Wang,Stavros Melemenidis,Rakesh Manjappa,Vignesh Viswanathan,Ramish M. Ashraf,Karen Levy,Lawrie B. Skinner,Luis A. Soto,Stephanie Chow,Brianna Lau,Ryan B. Ko,Edward E. Graves,Amy S. Yu,Karl K. Bush,Murat Surucu,Erinn B. Rankin,Billy W. Loo,Emil Schüler,Peter G. Maxim
DOI: https://doi.org/10.1002/mp.17432
IF: 4.506
2024-09-29
Medical Physics
Abstract:Background FLASH radiation therapy (RT) offers a promising avenue for the broadening of the therapeutic index. However, to leverage the full potential of FLASH in the clinical setting, an improved understanding of the biological principles involved is critical. This requires the availability of specialized equipment optimized for the delivery of conventional (CONV) and ultra‐high dose rate (UHDR) irradiation for preclinical studies. One method to conduct such preclinical radiobiological research involves adapting a clinical linear accelerator configured to deliver both CONV and UHDR irradiation. Purpose We characterized the dosimetric properties of a clinical linear accelerator configured to deliver ultra‐high dose rate irradiation to two anatomic sites in mice and for cell‐culture FLASH radiobiology experiments. Methods Delivered doses of UHDR electron beams were controlled by a microcontroller and relay interfaced with the respiratory gating system. We also produced beam collimators with indexed stereotactic mouse positioning devices to provide anatomically specific preclinical treatments. Treatment delivery was monitored directly with an ionization chamber, and charge measurements were correlated with radiochromic film measurements at the entry surface of the mice. The setup for conventional dose rate irradiation utilized the same collimation system but at increased source‐to‐surface distance. Monte Carlo simulations and film dosimetry were used to characterize beam properties and dose distributions. Results The mean electron beam energies before the flattening filter were 18.8 MeV (UHDR) and 17.7 MeV (CONV), with corresponding values at the mouse surface of 17.2 and 16.2 MeV. The charges measured with an external ion chamber were linearly correlated with the mouse entrance dose. The use of relay gating for pulse control initially led to a delivery failure rate of 20% (± 1 pulse); adjustments to account for the linac latency improved this rate to 95% agreement. Conclusion We successfully configured a clinical linear accelerator for increased output and developed a robust preclinical platform for anatomically specific irradiation, with highly accurate and precise temporal and spatial dose delivery, for both CONV and UHDR irradiation applications.
radiology, nuclear medicine & medical imaging