Combined generator/separator for continuous hydride generation: application to on-line pre-reduction of arsenic(V) and determination of arsenic in water by atomic emission spectrometry

Ian D. Brindle,Hosen Alarabi,Samir Karshman,Xiau-chun Le,Shaoguang Zheng,Hengwu Chen
DOI: https://doi.org/10.1039/an9921700407
1992-01-01
The Analyst
Abstract:A continuous hydride generator is described which incorporates a combined generator/separator. The low acid concentrations necessary in the tetrahydroborate(III)—L-cystein mediated determination of most hydride-forming elements require that the gas is purged from solution. As the hydride transfer reaction is so rapid, this allows both the hydride-forming reaction and the stripping reaction to take place in the same vessel. The system has been further modified to allow on-line pre-reduction of arsenic(V) to arsenic(III) by L-cysteine. Detection limits for arsenic, germanium and selenium are 4, 0.3 and 8 ng ml–1, respectively, at a sample flow rate of 15 ml min–1. Application of the system to the determination of arsenic in water is described. Concentrations of arsenic in several Ontario waters were found to be in the range 7–600 ng ml–1, Interferences from nickel and palladium, two of the most strongly interfering elements in the continuous hydride generation process, are diminished to a greater extent with the system described here than with the conventional high acid continuous system.