Experimental Consequences of Differences Between Composition of the Sample Solvent and of the Mobile Phase in HPLC

B. Porsch
DOI: https://doi.org/10.1080/10826079308019697
1993-10-01
Journal of Liquid Chromatography
Abstract:Anomalous peaks may be expected to occur even in cases when sample solubility in mobile phase is sufficient if sample solvent and mobile phase differ considerably in viscosity and/or elution power of sample solvent exceeds mobile phase strength substantially. Then, to improve peak shapes, the sample solvent—mobile phase viscosity ratio should be kept fairly below two and too high elution power of the sample solvent should be decreased by mixing with mobile phase prior to injection.