Comparison of Accuracy for HKA by X-ray and Knee Motion Analysis System and the Relationships between HKA and Gait Posture

Hui Zhang,Yanan Chen,Huiquan Jiang,Wenqing Yan,Yuanming Ouyang,Wei Wang,Yaru Liu,Ying Zhou,Shiyi Gu,Hong Wan,Axiang He,Yanjie Mao,Wanjun Liu
DOI: https://doi.org/10.21203/rs.3.rs-798002/v1
2021-08-20
Abstract:Abstract This study is aimed at obtaining the accuracy of the lower limb mechanical axis (Hip-Knee-Ankle, HKA) by the six degrees of freedom (6DOF) of the knee and ground gait data and analyzing the correlation between the lower limb mechanical axis and 6DOF. We analyzed 3D knee kinematics during ground gait of 99 patients with KOA and 80 patients 6 months after the operations with the vivo infrared navigation 3D portable knee joint movement analysis system (Opti-Knee®, Innomotion Inc, Shanghai, China). Significant correlations with low coefficients (r = -0.19, p = 0.01) between HKA value and anterior-posterior displacement for the whole cohort were found. There were significant correlations with moderate to high coefficients (r = 0.784 to 0.976) between the comparisons of HKA measured on the full-length alignment radiographs and 3D knee joint movement analysis system (Opti-Knee). The further linear correlation analysis showed that there was a significant correlation between the values of HKA measured by X-ray and movement analysis system(R 2 = 0.90, p < 0.01). Data with equivalent results as HKA could be provided by knee joint movement analysis system comparing with the conventional X-rays. Nonetheless, there was low significance between the HKA and ground gait data .