A sub-millimetric, 0.25 mN resolution fully integrated fiber-optic force-sensing tool for retinal microsurgery

Iulian Iordachita,Zhenglong Sun,Marcin Balicki,Jin U. Kang,Soo Jay Phee,James Handa,Peter Gehlbach,Russell Taylor
DOI: https://doi.org/10.1007/s11548-009-0301-6
2009-04-15
International Journal of Computer Assisted Radiology and Surgery
Abstract:PurposeRetinal microsurgery requires extremely delicate manipulation of retinal tissue where tool-to-tissue interaction forces are usually below the threshold of human perception. Creating a force-sensing surgical instrument that measures the forces directly at the tool tip poses great challenges due to the interactions between the tool shaft and the sclerotomy opening.MethodsWe present the design and analysis of a force measurement device that senses distal forces interior to the sclera using 1-cm long, 160 μm diameter Fiber Bragg Grating (FBG) strain sensors embedded in a 0.5 mm diameter tool shaft. Additionally, we provide an algorithm developed to cancel the influence of environmental temperature fluctuations.ResultsThe force-sensing prototype measures forces with a resolution of 0.25 mN in 2 DOF while being insensitive to temperature.ConclusionSub-millinewton resolution force sensors integrated into microsurgical instruments are feasible and have potential applications in both robotic and freehand microsurgery.
engineering, biomedical,radiology, nuclear medicine & medical imaging,surgery
What problem does this paper attempt to address?