Implantable multi sensor system for in vivo monitoring of cardiovascular parameters

P. Bingger,J. Fiala,A. Seifert,N. Weber,A. Moser,F. Goldschmidtboeing,K. Foerster,C. Heilmann,F. Beyersdorf,P. Woias,H. Zappe
DOI: https://doi.org/10.1109/sensor.2009.5285821
2009-06-01
Abstract:A novel implantable sensor for continuous long-term monitoring of various cardiovascular parameters is presented. An optical arterial blood oxygen saturation sensor and a piezoelectric blood pressure sensor have been developed using a silicone-based fabrication technology. An elastic silicone stripe houses the sensors and is wrapped around an arterial blood vessel. Due to this choice of a soft material, no necking of the vessel or influence on the blood flow occur even at large dilatations of 10 %. The pulse oximetric sensor is able to measure oxygen saturations in the physiologically relevant range of 65 %-100 % ($psO_{2}$) with a precision of ±1 %. The piezoelectric blood pressure sensor uses a 70 μm cellular polypropylene (PP) layer which responds to the dilatation of the blood vessel with a sensitivity of 0.75 mV/mmHg.