Antibacterial efficacy and cytotoxicity of low intensity direct current activated silver–titanium implant system prototype

Zhuo Tan,Edward A. Havell,Paul E. Orndorff,Rohan A. Shirwaiker
DOI: https://doi.org/10.1007/s10534-017-9993-1
2017-01-16
BioMetals
Abstract:Silver-based devices activated by electric current are of interest in biomedicine because of their broad-spectrum antimicrobial activity. This study investigates the in vitro antibacterial efficacy and cytotoxicity of a low intensity direct current (LIDC)-activated silver–titanium implant system prototype designed for localized generation and delivery of silver ions at the implantation site. First, the antibacterial efficacy of the system was assessed against methicillin-resistant Staphylococcus aureus (MRSA) over 48 h at current levels of 3 and 6 µA in Mueller–Hinton broth. The cytotoxicity of the system was then evaluated over 48 h in two phases using an in vitro model with in which the activated electrodes were suspended in growth medium in a cell-seeded tissue culture plate. In phase-1, the system was tested on human osteosarcoma (MG-63) cell line and compared to titanium controls. In phase-2, the cytotoxicity characteristics were validated with normal human diploid osteoblast cells. The LIDC-activated system demonstrated high antimicrobial efficacy against MRSA, but was also toxic to human cells immediately surrounding the electrodes. The statistical analysis showed that the cytotoxicity was a result of the presence of silver, and the electric activation did not make it worse.
biochemistry & molecular biology
What problem does this paper attempt to address?