Abstract:Background Molecular testing is increasingly needed in colorectal carcinoma (CRC) and the current clinically relevant mutations are in BRAF , KRAS and NRAS . This study aimed to further validate a new alternative polymerase chain reaction (PCR) platform (Idylla, Biocartis) against existing next-generation sequencing (NGS) and immunohistochemistry (IHC) assays. Methods 56 Idylla tests were performed on 43 CRC cases, in a total of 74 comparisons against an NGS panel (Ion Torrent) and the VE1 (anti- BRAF ) antibody IHC. Discrepant cases were also compared with either conventional (Cobas) or droplet digital PCR (Bio-Rad). Results Idylla showed an overall concordance of 100% (95% CI 93% to 100%) with comparator molecular testing and indications were that Idylla is likely to be more sensitive than routine NGS. BRAF IHC showed 90% concordance with NGS (95% CI 70% to 97%). Conclusions This study validates Idylla in formalin-fixed, paraffin-embedded CRC tissue. BRAF IHC, however, is an unreliable substitute for molecular testing in CRC.
pathology
What problem does this paper attempt to address?