Electrochemical DNA Biosensor Based on Graphene Oxide- Chitosan Hybrid Nanocomposites for Detection of Escherichia Coli O157:H7

Shichao Xu,
DOI: https://doi.org/10.20964/2017.04.16
IF: 1.541
2017-04-01
International Journal of Electrochemical Science
Abstract:In this article, we present a simple and sensitive electrochemical DNA biosensor based on graphene oxide (GO)/chitosan (CS) hybrid nanocomposites modified glassy carbon electrode (GCE) for detection of Escherichia coli O157:H7 (E.coli O157:H7). The morphology and composition of GO and hybrid nanocomposites were characterized by transmission electron microscope (TEM), X-ray powder diffraction (XRD), field emission scanning electron microscopy (FESEM), and Fourier transform infrared spectrum (FTIR). Cyclic voltammetry investigations indicated that the GO/CS/GCE showed excellent electron transfer ability and good linear relation. Under the optimal hybridization conditions, electrochemical impedance spectroscopy (EIS) responses of ssDNA/GO/CS/GCE biosensor were in linear with the target DNA in the concentration range from 1×10-14 to 1×10-8 M with the detection limit as 3.584×10-15 M (3σ). Moreover, differential pulse voltammetry (DPV) studies revealed good specificity and excellent ability of ssDNA/GO/CS/GCE biosensor to distinguish complementary, 1-base mismatched DNA, 2-base mismatched DNA and multi-base mismatched DNA. The developed strategy in this research revealed that the GO/CS modified electrode possess excellent performance for detecting of Escherichia coli O157:H7 DNA.
What problem does this paper attempt to address?