Impact of model-based iterative reconstruction on low-contrast lesion detection and image quality in abdominal CT: a 12-reader-based comparative phantom study with filtered back projection at different tube voltages

André Euler,Bram Stieltjes,Zsolt Szucs-Farkas,Reto Eichenberger,Clemens Reisinger,Anna Hirschmann,Caroline Zaehringer,Achim Kircher,Matthias Streif,Sabine Bucher,David Buergler,Luigia D’Errico,Sebastién Kopp,Markus Wilhelm,Sebastian T. Schindera
DOI: https://doi.org/10.1007/s00330-017-4825-9
IF: 7.034
2017-04-03
European Radiology
Abstract:ObjectivesTo evaluate the impact of model-based iterative reconstruction (MBIR) on image quality and low-contrast lesion detection compared with filtered back projection (FBP) in abdominal computed tomography (CT) of simulated medium and large patients at different tube voltages.MethodsA phantom with 45 hypoattenuating lesions was placed in two water containers and scanned at 70, 80, 100, and 120 kVp. The 120-kVp protocol served as reference, and the volume CT dose index (CTDIvol) was kept constant for all protocols. The datasets were reconstructed with MBIR and FBP. Image noise and contrast-to-noise-ratio (CNR) were assessed. Low-contrast lesion detectability was evaluated by 12 radiologists.ResultsMBIR decreased the image noise by 24% and 27%, and increased the CNR by 30% and 29% for the medium and large phantoms, respectively. Lower tube voltages increased the CNR by 58%, 46%, and 16% at 70, 80, and 100 kVp, respectively, compared with 120 kVp in the medium phantom and by 9%, 18% and 12% in the large phantom. No significant difference in lesion detection rate was observed (medium: 79-82%; large: 57-65%; P > 0.37).ConclusionsAlthough MBIR improved quantitative image quality compared with FBP, it did not result in increased low-contrast lesion detection in abdominal CT at different tube voltages in simulated medium and large patients.Key Points• MBIR improved quantitative image quality but not lesion detection compared with FBP.• Increased CNR by low tube voltages did not improve lesion detection.• Changes in image noise and CNR do not directly influence diagnostic accuracy.
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?