Effects of Ethanol Locks on Burst Pressure for Small Caliber Nonpower Injectable Peripherally Inserted Central Catheters

Mark D. Weber,Adam S. Himebauch,Jeremy C. Zuckerberg,Karla Resendiz,Thomas Conlon
DOI: https://doi.org/10.1097/pcc.0000000000003604
2024-11-05
Pediatric Critical Care Medicine
Abstract:In our previous work in which patients were consented for the use of ethanol lock therapy (ELT) for catheter salvage, we found the practice to be safe and effective in a wide range of polyurethane catheters (PURs) in the PICU ( 1 ). Within that series, we identified that two of eight small caliber, 2.6-Fr nonpower injectable peripherally inserted central catheters (PICCs) failed prematurely after ethanol exposure. It was unclear if the PICC failure was related to ethanol exposure or other factors. Ethanol exposure testing has not been performed by the manufacturer in these catheters and is considered off-label use. In this small bench study, we assessed the burst pressure of 2.6-Fr PICCs following various times of ethanol exposure to determine if ELT contributed to compromise the PUR PICC material.
pediatrics,critical care medicine