The effect of peripherally inserted central catheter (PICC) valve technology on catheter occlusion rates--the 'ELeCTRiC' study

Andrew J Johnston,Carmel T Streater,Remy Noorani,Joanne L Crofts,Aldwin B Del Mundo,Richard A Parker
DOI: https://doi.org/10.5301/jva.5000071
Abstract:Purpose: Peripherally Inserted Central Catheters (PICCs) are increasingly being used to provide short to medium-term central venous access. The current study was designed to test the hypothesis that PICC valve technology does not influence PICC occlusion rates. Methods: Intensive care unit (ICU) patients who required a PICC were randomized to one of three types of dual lumen PICC (open ended non-valved, Groshong valve, PASV valve). PICC occlusions were recorded and managed with a protocol that used urokinase. Results: A total of 102 patients were recruited to the study. The overall risk of occlusion per catheter was 35% (95% CI 26% to 44%). The overall rate of occlusion was 76 occlusions per 1000 catheter days (95% CI 61 to 95). Presence or type of valve did not significantly influence this rate (open-ended non-valved PICC 38% of catheters, 79 occlusions per 1000 catheter days; Groshong 38% of catheters, 60 occlusions per 1000 catheter days; PASV 27% of catheters, 99 occlusions per 1000 catheter days). The dose of urokinase required to treat PICC occlusions did not significantly differ between PICC types. Conclusions: Valved PICCs do not appear to influence PICC occlusion rates.
What problem does this paper attempt to address?