Third Generation Gamma Camera SPECT System

Narayan Bhusal
DOI: https://doi.org/10.48550/arXiv.2002.10707
IF: 4.506
2020-02-25
Medical Physics
Abstract:Single Photon Emission Computed Tomography (SPECT) is a non-invasive imaging modality, frequently used in myocardial perfusion imaging. The biggest challenges facing the majority of clinical SPECT systems are low sensitivity, poor resolution, and the relatively high radiation dose to the patient. New generation systems (GE Discovery, DSPECT) dedicated to cardiac imaging improve sensitivity by a factor of 5-8. The purpose of this work is to investigate a new gamma camera design with 21 hemi-ellipsoid detectors each with a pinhole collimator for Cardiac SPECT for further improvement in sensitivity, resolution, imaging time, and radiation dose. To evaluate the resolution of our hemi-ellipsoid system, GATE Monte-Carlo simulations were performed on point-sources, rod-sources, and NCAT phantoms. The purpose of point-source simulation is to obtain operating pinhole diameter by comparing the average FWHM (Full width half-maximum) of flat-detector system with curved hemi-ellipsoid detector system. The operating pinhole diameter for the curved hemi-ellipsoid detector was found to be 8.68mm. System resolution is evaluated using reconstructed rod-sources equally spaced within the region of interest. The results were compared with results of GE discovery system available in the literature. The system performance was also evaluated using the mathematical anthropomorphic NCAT (NURBSbased Cardiac Torso) phantom with a full (clinical) dose acquisition (25mCi) for 2 mins and an ultra-low-dose acquisition of 3mCi for 5.44mins.