An Off-Chip Capacitance Biosensor Based on Improved Cole-Cole Model for the Detection of Trimethylamine N-Oxide in Early Cardiovascular Disease

Wei-Cheng Lin,Chi-Jen Lo,Mei-Ling Cheng,Ren-Jie Zeng,Yun-Yu Hsieh,Peng-Ru Ho,Chun-Wei Yeh,Chun-Ting Hsieh,Ming-Chiu Chang,Chien-Hung Liao
DOI: https://doi.org/10.1109/jsen.2024.3368409
IF: 4.3
2024-01-01
IEEE Sensors Journal
Abstract:In this work, we demonstrated an off-chip capacitance array biosensor with limit of detection 1 μM of TMAO to diagnose chronic metabolism disease in urine. Based on the improved cole-cole model, the parameters of R catalyzed and C catalyzed and Rp catalyzed can predict the catalytic resistance of enzyme and redox effects of the analyte, and identify the AC properties of ion strength due to enzymatic catalysis. With the investigation of the off-chip capacitance biosensor electrical field change with respect to pixel geometry and associated driving and sensing electrode width in a pixel, the proposed off-chip biosensor with readout SoC exhibits high sensitivity of 21 ADC counts/μM (or 25 mV/μM), response time of 0.1 s, repetition of 98.9%, and drift over time of 0.5 mV. The proposed off-chip biosensor can distinguish TMAO in phosphate buffered saline (PBS) solution from the tiny capacitance change of TMAO induced by enzyme TorA with a difference of 2.15%, and the measurements has been successful verified with traditional cyclic-voltammetry (CV) method, which shows a difference of only 0.024%. The off-chip biosensor is designed to specifically detect TMAO by excluding the redox reactions of TMAO-specific enzyme TorA with creatine and creatinine in urine. This exclusion ensures that the biosensor focuses solely on TMAO, enhancing its specificity for accurate and reliable TMAO detection.
engineering, electrical & electronic,instruments & instrumentation,physics, applied
What problem does this paper attempt to address?