Impact of extracorporeal blood pump gap sizes on the performance and hemocompatibility under off‐design operation

Lars Fischer,Sebastian V. Jansen,Ulrich Steinseifer,Ifan Yen,Po‐Lin Hsu,Michael Neidlin
DOI: https://doi.org/10.1111/aor.14875
2024-09-28
Artificial Organs
Abstract:Our studies' results showed a modest sensitivity of gap size variations on key metrics and revealed a significant contribution of the gap regions to the device's total hemolytic footprint, with an increasing contribution for reduced flow rates. Furthermore, a trend toward larger uniformly sized gaps being beneficial regarding hemocompatibility was identified. Background Hemocompatibility remains the dominant challenge in rotary blood pumps, and more information on the relationship between individual pump design features, hemodynamics, and blood trauma in various operation conditions is necessary. The study evaluated the variation of gap sizes in extracorporeal blood pumps concerning their influence on blood compatibility, particularly during off‐design conditions. Methods We developed a parametric generic blood pump framework for in‐silico and in‐vitro design feature analysis. Thirty‐six designs with varying axial and radial gap sizes between 0.5 mm and 3 mm were generated. CFD was applied to calculate and compare device hemodynamics and evaluate the performance and hemocompatibility during off‐design and target operation conditions. The following quantities were analyzed: pressure difference, hemolysis potential, residence times, hydraulic efficiency, and recirculation ratio. Results The in‐vitro prototype showed excellent agreement with in‐silico predictions regarding hydraulic performance (R2 = 0.996 with a RMSE = 2.07). Our results show a modest impact of gap size variations ±10% on key metrics. Domain‐resolved analyses revealed a significant contribution of the gap regions to the device's overall hemolytic performance, with an increasing contribution for off‐design flow rates. Overall elevated hemolysis levels were identified if at least one gap size was held minimal. Conclusions We introduced and showed the feasibility of a parametric rotary blood pump framework to systematically investigate design feature impact. Results suggest, larger and uniformly sized gaps being overall beneficial regarding hemocompatibility.
engineering, biomedical,transplantation
What problem does this paper attempt to address?