On-orbit calibration of soft X-ray detector on Chang'E-2 satellite

Hong Xiao,Wen-Xi Peng,Huan-Yu Wang,Xing-Zhu Cui,Dong-Ya Guo
DOI: https://doi.org/10.1088/1674-1137/39/10/106002
IF: 2.944
2015-10-01
Chinese Physics C
Abstract:The X-ray spectrometer is one of the satellite payloads on the Chang'E-2 satellite. The soft X-ray detector is one of the devices on the X-ray spectrometer, designed to detect the major rock-forming elements within the 0.5–10 keV range on the lunar surface. In this paper, energy linearity and energy resolution calibration is done using a weak 55Fe source. Temperature and time effects are found not to give a large error. The total uncertainty of calibration is estimated to be within 5% after correction.
physics, particles & fields, nuclear