Suppression of Low Frequency Magnetic Drift Based on Magnetic Field Sensitivity in K-Rb-21Ne Atomic Spin Comagnetometer

Ye Liu,Wenfeng Fan,Yang Fu,Haoying Pang,Hongyu Pei,Wei Quan
DOI: https://doi.org/10.1109/tim.2022.3169541
IF: 5.6
2022-01-01
IEEE Transactions on Instrumentation and Measurement
Abstract:A spin-exchange relaxation-free comagnetometer is a promising high-precision gyroscope as its inertial measurement sensitivity can rival that of existing technologies. The magnetic field error is one of the main factors that limits the long-term stability of the comagnetometer. The steady-state response and frequency response models of a K-Rb-21Ne comagnetometer are modified by considering the spin-exchange rate between the electron spins and the nuclear spins. The modified models are used to redefine the bias magnetic sensitivity of the K-Rb-21Ne comagnetometer and the suppression method of low-frequency magnetic field error is proposed. We demonstrate the magnetic drift suppression method proposed in the article by optimizing the cell temperature and pump laser power density. This study is helpful to improve the rotation measurement performance of the K-Rb-21Ne comagnetometer.
engineering, electrical & electronic,instruments & instrumentation
What problem does this paper attempt to address?