A micro-electrochemical sensor based on a nanostructured carbon paste electrode as an original tool for selective detection of ultra-traces of 17α –ethinylestradiol in environmental water samples and in biologic fluids

Siham Amra,Saliha Bourouina-Bacha,Mustapha Bourouina,Didier Hauchard
DOI: https://doi.org/10.1016/j.microc.2024.110753
IF: 5.304
2024-05-20
Microchemical Journal
Abstract:A new electrochemical micro-sensor has been designed for the sensitive detection of ultra-traces of 17α – ethinylestradiol (EE 2 ). A carbon paste cavity micro-electrode modified by carbon black nanoparticles (GC-nCB)-μCPE was used with an electrochemical preconcentration (EP) step of EE 2 on (GC-nCB)-μCPE at 0.0 V during 300 s in a 0.01 mol.L -1 phosphate buffer solution (pH = 7.4) before voltammetric determination. The results demonstrate that two line calibrations were obtained in the concentration ranges of 0.005 to 0.2 μmol.L -1 and 0.2 to 1 μmol.L -1 . The detection and quantification limits are estimated at 1.4 nmol.L -1 and 4.9 nmol.L -1 , respectively. Moreover, a high sensitivity of 4.16 μA.mol −1 .L and good selectivity for the quantification of EE 2 in the presence of different interfering organic and inorganic compounds were observed. Based on the results (satisfactory recovery (less than 3.5 %) and relationship standard deviation (less than 2.5 %)), the designed micro-sensor can be successfully applied to the analysis of EE 2 in environmental water samples and in biologic fluids. The modified microsensor presented an efficient and inexpensive tool for the quantification of EE 2 .
chemistry, analytical
What problem does this paper attempt to address?