Green synthesized gold nanoparticles and CuO-based nonenzymatic sensor for saliva glucose monitoring

Younus Ali,Heman B. Abdulrahman,Wei-Ting Ting,Matiar M. R. Howlader,Md Younus Ali
DOI: https://doi.org/10.1039/d3ra05644a
IF: 4.036
2024-01-01
RSC Advances
Abstract:Gold nanoparticles (AuNPs) were synthesized using HAuCl 4 and orange peel extract. AuNPs and CuO modified screen printed carbon electrode (Au/CuO/SPCE) converts glucose to gluconolactone. This sensor was applied for detection of glucose in saliva.
chemistry, multidisciplinary
What problem does this paper attempt to address?
The problem this paper attempts to address is the development of a simple, low-cost, and highly sensitive non-enzymatic electrochemical glucose sensor for monitoring glucose in saliva. Specifically, the authors aim to utilize green-synthesized gold nanoparticles (AuNPs) and wet-chemically synthesized copper oxide (CuO) nanoparticles to construct a composite sensor (Au/CuO/SPCE) on a screen-printed carbon electrode (SPCE). This sensor can effectively detect low concentrations of glucose in saliva, providing a non-invasive monitoring method for the management of chronic diseases such as diabetes. ### Background and Problem 1. **Importance of Blood Glucose Monitoring**: - Glucose is crucial for brain and muscle function. - Abnormal blood glucose levels can lead to various fatal diseases, such as diabetes, hyperglycemia, and hypoglycemia. - Approximately 422 million adults worldwide have diabetes. - Regular blood glucose monitoring is essential for early detection, tracking disease progression, and evaluating treatment efficacy. 2. **Limitations of Existing Monitoring Methods**: - Blood glucose monitoring typically requires finger pricking, which is invasive and may cause discomfort and infection risk. - Sweat and interstitial fluid monitoring, while less invasive, can cause skin irritation and discomfort. - Enzymatic sensors are costly, have poor stability, and are easily affected by environmental conditions. 3. **Advantages of Non-Enzymatic Sensors**: - Non-enzymatic sensors can overcome the above issues, providing a low-cost, highly stable, and highly sensitive monitoring method. - The glucose concentration in saliva is relatively low, requiring highly sensitive sensors for effective monitoring. ### Solution The authors developed this non-enzymatic sensor through the following steps: 1. **Green Synthesis of Gold Nanoparticles (AuNPs)**: - Using orange peel extract (OPE) as a reducing and stabilizing agent to synthesize stable gold nanoparticles. - This method avoids the use of toxic chemicals, offering environmental benefits. 2. **Wet Chemical Synthesis of Copper Oxide (CuO) Nanoparticles**: - Synthesizing CuO nanoparticles using copper chloride (CuCl₂) and sodium hydroxide (NaOH). - Dropping CuO nanoparticles onto SPCE to form the CuO/SPCE electrode. 3. **Preparation of the Composite Sensor**: - Further dropping green-synthesized AuNPs onto the CuO/SPCE electrode to form the Au/CuO/SPCE composite sensor. - Optimizing parameters such as the amount of AuNPs and CuO, scan rate, and NaOH concentration to enhance sensor performance. ### Experimental Results - **Linear Response Range**: The sensor's linear response range for glucose is from 2 mM to 397 mM, with a sensitivity of 236.70 mAm M⁻¹cm⁻². - **Stability and Reproducibility**: The sensor exhibits good stability and reproducibility. - **Selectivity**: The sensor shows high selectivity for glucose in the presence of different biomolecules. - **Practical Application**: The effectiveness of the sensor was validated by measuring glucose in human saliva, demonstrating its potential for non-invasive glucose monitoring. ### Conclusion This study developed a non-enzymatic electrochemical glucose sensor based on green-synthesized AuNPs and CuO nanoparticles, featuring high sensitivity, good stability, and high selectivity. It is suitable for non-invasive monitoring of glucose in saliva, providing a new tool for the management of chronic diseases such as diabetes.