Validating QT-Interval Measurement Using the Apple Watch ECG to Enable Remote Monitoring During the COVID-19 Pandemic
Marc Strik,Théo Caillol,F. Daniel Ramirez,Saer Abu-Alrub,Hugo Marchand,Nicolas Welte,Philippe Ritter,Michel Haïssaguerre,Sylvain Ploux,Pierre Bordachar
DOI: https://doi.org/10.1161/CIRCULATIONAHA.120.048253
IF: 37.8
2020-07-29
Circulation
Abstract:Screening and monitoring for QT prolongation when certain medications are initiated are routinely performed to avoid arrhythmic complications. However, the novel coronavirus disease 2019 (COVID-19) pandemic and its proposed treatments—including hydroxychloroquine and azithromycin, which are known to prolong the QT interval 1 —raise logistical and safety concerns with established QT monitoring strategies. Recently, the American College of Cardiology made recommendations for QT monitoring in outpatients with COVID-19 on hydroxychloroquine/azithromycin, suggesting that the use of direct-to-consumer mobile devices such as the Apple Watch 1-lead ECG could be considered in cases of resource constraints or quarantines. 2 The Apple Watch ECG is cleared by the US Food and Drug Administration for detecting atrial fibrillation but has not been studied for QT monitoring. Lead I (the lead recorded by the Apple Watch) may be suboptimal for measuring this interval; however, other leads can be reproduced by placing the smartwatch on the left ankle or chest. 3 We therefore sought to validate the use of the Apple Watch for QT measurement, including using nonstandard smartwatch positions, in an unselected outpatient population. Between December 2019 and January 2020, 100 consecutive patients in sinus rhythm were enrolled from outpatient or emergency departments. The study was approved by our institutional review committee, and the subjects gave informed consent. Standard 12-lead ECGs were performed, followed by smartwatch electrocardiographic recordings using the Apple Watch Series 4 (Apple Inc, Cupertino, CA). After a brief demonstration, patients recorded 30-second Apple Watch electrocardiographic equivalents of lead I (AW-I; watch on left wrist), lead II (AW-II; watch on left ankle), and a simulated lead V 6 (AW-LAT; watch on left lateral chest; Figure [A ]). Using commercially available software (EP Calipers, EP studios Inc, Louisville, KY), a cardiologist measured 3 RR and QT intervals to calculate the corrected QT interval (QTc) using the Bazett formula ( Figure [B ]). A QTc >480 milliseconds was considered high risk. Agreement between the 12-lead and Apple Watch QTc measurements was calculated with the use of the median absolute error and Bland-Altman analyses. To measure interobserver variability, all AW-I QTc measurements were repeated by a second blinded cardiologist, and the intraclass correlation coefficient was calculated on the basis of a 2-way random absolute agreement model with single measurements. Agreement on whether a tracing was interpretable was assessed with the Cohen κ statistic. T-wave amplitude was measured to evaluate its association with QTc measurement accuracy. Download figure Download PowerPoint Figure. Apple Watch electrocardiographic recordings, QT interval measurements, and impact of recording tracings from nonstandard smartwatch positions. A , Recording Apple Watch electrocardiographic equivalents of lead I (AW-I; watch on left wrist), lead II (AW-II; watch on left ankle), and a simulated lead V 6 (AW-LAT; watch on left lateral chest). B , Corrected QT interval (QTc) measurements from a 12-lead ECG and an Apple Watch lead I ECG (AW-I). QTc using the 12-lead ECG is 372 milliseconds; QTc using AW-I is 380 milliseconds (absolute difference, 8 milliseconds; bias, −8 milliseconds). C , Repositioning the Apple Watch may improve signal quality and measurement of the QT interval. Example of 2 smartwatch electrocardiographic recordings from the same patient. Measurement of the QT interval was deemed impossible in lead II (AW-II) because of low T-wave amplitude (≤0.1 mV). After the smartwatch was moved to the left lateral chest wall (AW-LAT), the T-wave amplitude increased to 2 mV, markedly facilitating the measurement of the QT interval. Note the similar increase in QRS complex amplitudes. The mean age was 67±7 years; 59% were male; and 35% had diagnosed cardiac disease. Heart rates were similar on 12-lead and Apple Watch recordings (69±11 bpm versus 70±13 bpm, respectively; P =0.2, paired Student t test). QTc intervals ranged from 336 to 530 milliseconds on 12-lead ECGs. Eight patients were identified as high risk, all of whom were similarly identified by the smartwatch. QTc measurements could not be -Abstract Truncated-
cardiac & cardiovascular systems,peripheral vascular disease