A comparison of three different microwave systems for laparoscopic liver tumor ablation

Eren Berber,Ali Gunduz Sarioglu,Ege Akgun
DOI: https://doi.org/10.1002/jso.27621
2024-03-14
Journal of Surgical Oncology
Abstract:Background Our aim was to perform a comparison of three current microwave ablation (MWA) systems widely used for laparoscopic liver ablations in terms of ablation kinetics and geometry of ablation zones. Methods This was a retrospective, institutional review board‐approved study comparing Emprint, Emprint HP, and NeuWave systems for laparoscopic liver ablation. Analyses were performed via Mann−Whitney U and χ2 tests. Continuous data are presented as median (interquartile range). Results For Emprint, Emprint HP, and NeuWave groups, tumor size was 1.16 (0.8), 1.21 (0.7), and 1.27 (0.9) cm (p = 0.54). Ablation time per lesion was 7 (6), 4 (2.8), and 4 (3.3) min (p
oncology,surgery
What problem does this paper attempt to address?