Using Patient-Specific 3D Modeling and Simulations to Optimize Microwave Ablation Therapy for Liver Cancer

Amirreza Heshmat,Caleb S. O'Connor,Jessica Albuquerque Marques Silva,Iwan Paolucci,Aaron Kyle Jones,Bruno C. Odisio,Kristy K. Brock
DOI: https://doi.org/10.3390/cancers16112095
2024-05-31
Cancers
Abstract:Microwave ablation (MWA) of liver tumors presents challenges like under- and over-ablation, potentially leading to inadequate tumor destruction and damage to healthy tissue. This study aims to develop personalized three-dimensional (3D) models to simulate MWA for liver tumors, incorporating patient-specific characteristics. The primary objective is to validate the predicted ablation zones compared to clinical outcomes, offering insights into MWA before therapy to facilitate accurate treatment planning. Contrast-enhanced CT images from three patients were used to create 3D models. The simulations used coupled electromagnetic wave propagation and bioheat transfer to estimate the temperature distribution, predicting tumor destruction and ablation margins. The findings indicate that prolonged ablation does not significantly improve tumor destruction once an adequate margin is achieved, although it increases tissue damage. There was a substantial overlap between the clinical ablation zones and the predicted ablation zones. For patient 1, the Dice score was 0.73, indicating high accuracy, with a sensitivity of 0.72 and a specificity of 0.76. For patient 2, the Dice score was 0.86, with a sensitivity of 0.79 and a specificity of 0.96. For patient 3, the Dice score was 0.8, with a sensitivity of 0.85 and a specificity of 0.74. Patient-specific 3D models demonstrate potential in accurately predicting ablation zones and optimizing MWA treatment strategies.
oncology
What problem does this paper attempt to address?
The problem that this paper attempts to solve is the challenges existing in the process of microwave ablation (MWA) for liver cancer treatment, especially how to optimize the MWA treatment strategy through patient - specific three - dimensional modeling and simulation to improve the treatment effect and reduce the damage to healthy tissues. Specifically, the research aims to: 1. **Develop personalized 3D models**: Utilize the patient's CT image data to construct personalized 3D models to simulate the temperature distribution and tumor destruction during the MWA process. 2. **Predict the ablation area**: Verify the consistency between the predicted ablation area and the clinical results through simulation, thereby providing accurate treatment planning before treatment. 3. **Optimize treatment parameters**: Determine the optimal treatment time, power input and other parameters to ensure that the minimum ablation margin (MAM) of at least 5 mm is achieved while minimizing the damage to the surrounding healthy tissues as much as possible. ### Background and challenges - **Advantages of MWA**: MWA is widely used in the treatment of liver cancer because of its efficient heating characteristics, low sensitivity to the heat sink effect, short ablation time and large ablation area. - **Existing challenges**: Although MWA has many advantages, achieving effective ablation (defined as a minimum ablation margin of at least 5 mm) still faces challenges. These challenges include the influence of factors such as tumor size and location, antenna placement, patient health status and MWA device parameters. - **Limitations of manual adjustment**: Antenna placement usually requires manual adjustment and precise monitoring during the treatment process, which often leads to multiple adjustments and potential over - ablation, which may cause irreversible damage to healthy tissues. ### Research methods - **Patient selection**: Three patients who received MWA treatment were selected. These patients had complete data and were suitable for 3D modeling. - **3D model construction**: Use the patient's CT image data to extract the surface data of the liver, tumor and blood vessels through segmentation technology and convert them into 3D volume models. - **Simulation method**: Combine the coupled equations of electromagnetic wave propagation and bio - heat transfer, and use the finite element method (FEM) to simulate in COMSOL Multiphysics to predict the temperature distribution and tissue damage. ### Main findings - **Optimization of ablation time**: For patient 1, complete tumor ablation was achieved at 8 minutes, and continuing the treatment until 10 minutes would only increase the damage to healthy tissues. For patient 2 and patient 3, complete tumor ablation was achieved at 8 minutes and 7 minutes respectively. Early termination of treatment can significantly reduce the damage to healthy tissues. - **Temperature distribution analysis**: The simulation results show that when the temperature in the tumor exceeds 50 °C, cancer cells can be effectively destroyed. Although an overly long ablation time will expand the ablation area, it will not significantly improve the effect of tumor destruction, but will increase the damage to healthy tissues. ### Conclusion Through patient - specific 3D modeling and simulation, the ablation area of MWA can be accurately predicted, treatment parameters can be optimized, and unnecessary damage to healthy tissues can be reduced, thereby improving the treatment effect and the prognosis of patients.