The evaluation of a non-invasive respiratory monitor in ards patients in supine and prone position

Tommaso Pozzi,Silvia Coppola,Elena Chiodaroli,Federico Cucinotta,Francesca Becci,Davide Chiumello
DOI: https://doi.org/10.1007/s10877-024-01147-0
IF: 1.977
2024-03-28
Journal of Clinical Monitoring and Computing
Abstract:Purpose: The Prone positioning in addition to non invasive respiratory support is commonly used in patients with acute respiratory failure. The aim of this study was to assess the accuracy of an impedance-based non-invasive respiratory volume monitor (RVM) in supine and in prone position. Methods: In sedated, paralyzed and mechanically ventilated patients in volume-controlled mode with acute respiratory distress syndrome scheduled for prone positioning it was measured and compared non-invasively tidal volume and respiratory rate provided by the RVM in supine and, subsequently, in prone position, by maintaining unchanged the ventilatory setting. Results: Forty patients were enrolled. No significant difference was found between measurements in supine and in prone position either for tidal volume ( p = 0.795; p = 0.302) nor for respiratory rate ( p = 0.181; p = 0.604). Comparing supine vs. prone position, the bias and limits of agreements for respiratory rate were 0.12 bpm (-1.4 to 1.6) and 20 mL (-80 to 120) for tidal volume. Conclusions: The RVM is accurate in assessing tidal volume and respiratory rate in prone compared to supine position. Therefore, the RVM could be applied in non-intubated patients with acute respiratory failure receiving prone positioning to monitor respiratory function.